Distal Axial Pin Lock without or with Nail Lock of Femoral Hybrid Intramedullary Nailing (FHIN) for Femoral Shaft and Distal Femoral Fractures - A Novel Technique of Distal Lock

Authors

  • M. Amin Cheema Hammad Amin Cheema Senior Consultant, Millat Orthopaedic Hospital, Sargodha

Keywords:

Distal axial pin lock, femoral hybrid intramedullary nailing; femoral shaft and distal femoral fractures.

Abstract

Objective: A new technique of distal axial pin lock without or with nail lock for distal locking of femoral hybrid intramedullary nailing (FHIN) was devised for fixation of femoral shaft and distal femoral fractures. It could redress the difficulties of distal locking, hazards of extensive radiation exposure in lateral projection and the susceptibilities of screw/nail bend/break and mediolateral toggling of distal end of nail after conventional locking of femoral locked intramedullary nails (FLINS).

Methods: 63 femoral shaft and distal femur fractures, treated by FHIN were prospectively assessed in this study. All FHIN implantations were distally locked with axial pin without or with nail lock depending on level and comminution of fracture. Axially stable femoral shaft fractures (FSFs) were distal locked with axial pin only. Axially unstable FSFs were distal locked with axial pin and oval nail hole. While distal femoral fractures were locked with axial pin and both oval and round nail holes. Postoperative assessment was made for accuracy of distal axial pin locking procedure. Technical distal axial pin lock problems and inability to find the nail slot were also recorded. Patients were followed until the time of solid union of fractures. During follow up X-rays patients were evaluated for complications of distal screw/nail bend /break and toggling of nail.

Results: Of a total 63 distal axial pin lock procedures the accuracy of installation was 100% with single attempts and significantly shortened distal axial pin locking time. During follow up complications of distal axial pin/screw/nail bend/break, toggling of distal end of nail, distal migration of the nail, and loss of reduction were not noted. The number of radiation exposures, distal locking time and total operation time was significantly reduced as image intensifier use and its lateral adjustment was not required. The technique was quite easier and straightforward with shorter time period and learning curve.

Conclusion: Primary distal axial pin lock without or with nail hole(s)lock of FHIN for treatment of femoral fractures can reduce the incidence of distal screw/nail bend/break and toggling of distal end of nail to the negligible extent. It also looks to be a predictable solution to decrease radiation dosage and its vulnerability during FHIN fixation by avoidance of image intensifier exposure and its lateral adjustment. Thus, reducing the radiation exposure and total surgical time.

References

Winquist RA, Hansen ST, Clawson DK. Closed intramedullary nailing of femoral fractures. J Bone Joint Surg [Am]. 1984; vol 66: 529-30.

Yong-qing W, Yong-cheng H, Zhan-min X, Ya-wei Z,Jia-min W: An intramedullary nail with multifunctional interlocking for all types of fracture in both femurs. Orthopaedic Surgery 2009; vol 1 (2): 121–126.

Pardiwala D, Prabhu V, Dudhniwala G, et al. The AO distal locking aiming device-an evaluation of efficacy and learning curve: Injury. 2001; vol 32: 713–718.

Kamarianakis Z, Buliev I, Pallikarakis N. Robust identi?cation and localization of intramedullary nail holes for distal locking using CBCT-a simulation study: Med Eng Phys. 2011; vol 33: 479–489.

Metsemakers WJ, Wijnen V, Sermon A, Vanderschot P, Nijs S. Intramedullary nailing of humeral shaft fractures-failure analysis of a single Centre series: Arch Orthop Trauma Surg. 2015; vol 135: 1391–1399.

Wang Y, Hu Y, Xu Z, Zhao Y,Wu J. An intramedullary nail with multifunctional interlocking for all types of fracture in both femurs: Orthopaedic Surgery. 2009; Vol 1, (2): 121–126.

Whatling GM, Nokes LD. Literature review of current techniques for the insertion of distal screws into intramedullary locking nails: Injury. 2006; vol 37(2):109–119.

Lin J, Inoue N, Valdevit A, Hang YS, Hou SM, Chao EYS. Biomechanical comparison of antegrade and retrograde nailing of humeral shaft fractures: Clin Orthop. 1998; vol 351:203-13.

Bucholz RW, Ross DSE, Lawrence KL. Fatigue fracture of the interlocking nail in the treatment of fractures of the distal part of the femoral shaft: J Bone Joint Surg [Am]. 1987; vol 69:1391-9.

Leloup T, El Kazzi W, Schuind F, Warzee N. A novel technique for distal locking of intramedullary nail based on two non-constrained fluoroscopic images and navigation: IEEE Trans Med Imaging. 2008; vol 27(9): 1202-1212.

Leroux T, A Khoshbin, MT Nousiainen. Training Distal Locking Screw Insertion Skills to Novice Trainees-A Comparison Between Fluoroscopic- and Electromagnetic-Guided Techniques: J Orthop Trauma. 2015; vol 29(10): 441-446.

Jinn L, Son-Jyh L, Po-Quang C, Shu-Hua Y. Stress analysis of the distal locking screws for femoral interlocking nailing: Journal of Orthopaedic Research. 200I; vol 19: 57-63.

Bucholz RW, Ross DSE, Lawrence KL. Fatigue fracture of the interlocking nail in the treatment of fractures of the distal part of the femoral shaft: J Bone Joint Surg [Am]. 1987; vol 69:1391-9.

Tencer AF, Johnson KD. Low extremity fixation. In Biomechanics in orthopedic trauma Bone fracture and fixation: Philadelphia JB Lippincott; 1994: 249-304.

Wu CC, Shih CH. Biomechanical analysis of the mechanism of interlocking nail failure: Arch Orthop Trauma Surg. 1992; vol 111: 268-72.

Durall I, Diaz-Bertrana MC, Puchol JL, Franch J. Radiographic findings related to interlocking nailing-windshield-wiper effect and locking screw failure: Vet. Comp. Orthop. Traumatol. 2003; vol 16:217-222.

Durall I, Falcón C, Diaz-Bertrana MC, Franch J. Effects of static fixation and dynamization after interlocking femoral nailing locked with an external fixator-An experimental study in dogs: Vet. Surg. 2004; vol 33:323-332.

Yong-qing W , Yong-cheng H, Zhan-min X , Ya-wei Z , Jia-min W. An intramedullary nail with multifunctional interlocking for all types of fracture in both femurs: Orthopedic Surgery. 2009; vol 1 (2): 121–126.

Whatling GM, Nokes LDM. Literature review of current techniques for the insertion of distal screws into intramedullary locking nails: Injury. 2006; vol 37:109–119.

Wang Y, Han B, Shi Z, et al. Comparison of free-hand fluoroscopic guidance and electromagnetic navigation in distal locking of tibia intramedullary nails: Medicine (Baltimore). 2018; vol 97(27).

Zhao X, Fan Y, Chen J. A comparison of free-hand method and electromagnetic navigation technique for the distal locking during intramedullary nailing procedures- a meta-analysis: Arch Orthop Trauma Surg. 2021; vol 141(1):45–53.

Downloads

Published

2023-10-11

How to Cite

Distal Axial Pin Lock without or with Nail Lock of Femoral Hybrid Intramedullary Nailing (FHIN) for Femoral Shaft and Distal Femoral Fractures - A Novel Technique of Distal Lock. (2023). Journal of Pakistan Orthopaedic Association, 35(03), 152-158. http://jpoa.org.pk/index.php/upload/article/view/800

Similar Articles

1-10 of 194

You may also start an advanced similarity search for this article.